Laser-based Elemental Analyzer

EA-300 series

Catalogues View Catalogue

Laser-based Elemental Analyzer EA-300

EA-300 - Laser-based Elemental Analyzer

*Please note that accessories depicted in the image are for illustrative purposes only and may not be included with the product.

Data Sheet (PDF)

  • CE Marking
  • CSA

Specifications

Model

EA-300

Analysis optical system

Detection principle

Laser-induced breakdown spectroscopy

Supported elements

H1 to U92

Laser

Laser type

Nd: YAG laser

Laser class

Class 1 laser product (IEC/EN 60825-1, JIS C 6802, FDA (CDRH) Part 1040.10)

Laser wavelength

355 nm

Spot size

10 μm (typ.)

Observation optical system

Image sensor

1/1.8-inch, 3.19-megapixel CMOS image sensor

Observation magnification

300× to 1000×

Scanning system

Progressive

Frame rate

50 fps (max.)

Optical system

Lens type

Specially designed reflective objective lens

Objective lens working distance

25 mm

High dynamic range

16-bit intensity range through RGB data from each pixel

Gain

Manual, Preset

Electronic shutter

Auto, Manual, 1/60, 1/120, 1/250, 1/500, 1/1000, 1/2000, 1/5000, 1/9000, 1/19000

Supercharge shutter

Can be set in 0.01 s increments from 0.02 to 16 s

White balance

Push-set, Manual, Preset (2700K, 3200K, 5600K, 9000K)

Built-in coaxial illumination light source

Type

High-brightness LED

Service life

100,000 hours (reference value)

Colour temperature

5810K (typ.)

Built-in ring illumination light source

Type

High-brightness LED

Service life

100,000 hours (reference value)

Colour temperature

5810K (typ.)

Power supply

Power supply method

Power supplied via dedicated cable from VHX Series controller camera port

Environmental resistance

Ambient temperature

10 to 33°C

Relative humidity

35 to 80% RH (No condensation)

Dimensions

280 (W) × 375 (H) × 210 (D) mm (with lens stored) (excluding protrusions)

Weight

Approx. 7.2 kg (Analysis system main unit)

Data Sheet (PDF) Other Models